mgh005 8 360/0.476 1706927 mgh005.dat 212 1267(-87)/mV 12 0 -727 27995 0 ECG lead I mgh005.dat 212 1254(-462)/mV 12 0 -967 62424 0 ECG lead II mgh005.dat 212 1057(-70)/mV 12 0 -834 53834 0 ECG lead V mgh005.dat 212 11.72(-1229)/mmHg 12 0 -1423 28509 0 ART mgh005.dat 212 20.02(-1020)/mmHg 12 0 -1279 29743 0 PAP mgh005.dat 212 19.56(-1033)/mmHg 12 0 -1141 2954 0 CVP mgh005.dat 212 1000 12 0 -837 31326 0 Resp. Imp. mgh005.dat 212 1000 12 0 -79 60930 0 CO2 #: 56 : M : Aortobifemoral graft; GI bleed; Exploratory laporotomy; Sepsis # PHARMACOLOGIC SUPPORT: # Dopamine @ 213 mcg/min # ELECTROCARDIOGRAPHIC DATA: # UNDERLYING RHYTHM: # Sinus tachycardia @ 100 bpm # ECG INTERPRETATION: # Old antero-lateral infarct # Anterior ST segment elevation consistent with aneurysm # HEMODYNAMIC DATA: # ART: 120/60 MEAN: 90 # PAP: 28/20 PCW: 14 (@ 26 min, 31 min, 42 min, 58 min) # RAP: 12 CO: 9.2 # WAVEFORM PATTERNS: # Respiratory variation # TECHNICAL COMMENTS: # ART pop-test unusual; tracing damped during recording # Transducer dome noted to be defective at end of case # RESPIRATORY DATA: # RATE: 12 bpm # MODE OF VENTILATION: # Controlled # CO2 RECORDING: # Sidestream sensor with Datex unit on @ 55 min